- Ultra-light body weight and ultra-light plunger force for max ergonomics
- 8 and 12 channel pipette are appropriate for 96-well plates
- Dispensing head rotating for effortless pipetting
- Individual piston and tip cone assemblies allow easy repair and maintenance
- Compound material-made tip cone securing high sealing performance
- Each pipette is supplied with calibration certificate according to ISO8655
- Compatible with most universal tip brands
- Fully autoclavable
Product features/Short Description:
- Ultra-light body weight and ultra-light plunger force for max ergonomics
- Available in adjustable or fixed volume, single or multi-channel
- Available volume ranges from 0.1µl to 10mL
- Easy calibration and maintenance; tool supplied
- Manufactured from new innovative materials
- Each pipette is supplied with calibration certificate according to ISO8655
- Fully autoclavable
Long Description
Waverly LitePette™ mechanical pipettes are newly designed with new innovative materials in order to be the most ergonomic pipette in the industry while maintaining durable chemical-resistant properties. LitePettes are Waverly’s most economically-priced pipettes for practical everyday lab use.
Available models include single-channel adjustable, single-channel fixed, and multichannel (8 & 12-channel) adjustable. They work with universal pipette tips from most manufacturers worldwide however we recommend using them with KaliTips; superior quality universal-fitting pipette tips that work perfectly with these and other pipettes.
The adjustable pipettes can be adjusted by turning the plunger knob in either direction to the desired volume which can be read in the display window on the side. Note: There is not a volume-locking function on these pipettes. (If you are seeking superior accuracy/precision, have a look at the Waverly Durapette™ series)
Fully autoclavable
Waverly LitePettes are fully autoclavable, withstanding steam sterilization at 121ºC for 20 minutes.
Calibration
All Waverly pipettes have been quality tested according to ISO8655-2:2002 and are supplied with individual calibration certificates. The quality control includes gravimetric testing of each pipette with distilled water at 22ºC.
What’s included
Each pipette comes with a calibration tool and one box of KaliTips.
Technical Data
8-channel Adjustable and Fixed volume
Volume Range |
Increment |
Test Volume |
Maximum permissible systematic error |
Maximum permissible random error |
||
(Inaccuracy) |
(Imprecision) |
|||||
% |
μL |
% |
μL |
|||
0.5-10μL |
0.1μL |
10μL |
1.50% |
0.15 |
1.50% |
0.15 |
5μL |
2.50% |
0.125 |
2.50% |
0.125 |
||
1μL |
4.00% |
0.04 |
4.00% |
0.04 |
||
5-50μL |
0.5μL |
50μL |
1.00% |
0.5 |
0.50% |
0.25 |
25μL |
1.50% |
0.375 |
1.00% |
0.25 |
||
5μL |
3.00% |
0.15 |
2.00% |
0.1 |
||
50-300μL |
5μL |
300μL |
0.70% |
2.1 |
0.25% |
0.75 |
150μL |
1.00% |
1.5 |
0.50% |
0.75 |
||
50μL |
1.50% |
0.75 |
0.80% |
0.4 |
12-channel Adjustable and Fixed volume
Volume Range |
Increment |
Test Volume |
Maximum permissible systematic error |
Maximum permissible random error |
||
(Inaccuracy) |
(Imprecision) |
|||||
% |
μL |
% |
μL |
|||
0.5-10μL |
0.1μL |
10μL |
1.50% |
0.15 |
1.50% |
0.15 |
5μL |
2.50% |
0.125 |
2.50% |
0.125 |
||
1μL |
4.00% |
0.04 |
4.00% |
0.04 |
||
5-50μL |
0.5μL |
50μL |
1.00% |
0.5 |
0.50% |
0.25 |
25μL |
1.50% |
0.375 |
1.00% |
0.25 |
||
5μL |
3.00% |
0.15 |
2.00% |
0.1 |
||
50-300μL |
5μL |
300μL |
0.70% |
2.1 |
0.25% |
0.75 |
150μL |
1.00% |
1.5 |
0.50% |
0.75 |
||
50μL |
1.50% |
0.75 |
0.80% |
0.4 |
8-channel Adjustable and Fixed volume
Volume Range |
Increment |
Test Volume |
Maximum permissible systematic error |
Maximum permissible random error |
||
(Inaccuracy) |
(Imprecision) |
|||||
% |
μL |
% |
μL |
|||
0.5-10μL |
0.1μL |
10μL |
1.50% |
0.15 |
1.50% |
0.15 |
5μL |
2.50% |
0.125 |
2.50% |
0.125 |
||
1μL |
4.00% |
0.04 |
4.00% |
0.04 |
||
5-50μL |
0.5μL |
50μL |
1.00% |
0.5 |
0.50% |
0.25 |
25μL |
1.50% |
0.375 |
1.00% |
0.25 |
||
5μL |
3.00% |
0.15 |
2.00% |
0.1 |
||
50-300μL |
5μL |
300μL |
0.70% |
2.1 |
0.25% |
0.75 |
150μL |
1.00% |
1.5 |
0.50% |
0.75 |
||
50μL |
1.50% |
0.75 |
0.80% |
0.4 |
12-channel Adjustable and Fixed volume
Volume Range |
Increment |
Test Volume |
Maximum permissible systematic error |
Maximum permissible random error |
||
(Inaccuracy) |
(Imprecision) |
|||||
% |
μL |
% |
μL |
|||
0.5-10μL |
0.1μL |
10μL |
1.50% |
0.15 |
1.50% |
0.15 |
5μL |
2.50% |
0.125 |
2.50% |
0.125 |
||
1μL |
4.00% |
0.04 |
4.00% |
0.04 |
||
5-50μL |
0.5μL |
50μL |
1.00% |
0.5 |
0.50% |
0.25 |
25μL |
1.50% |
0.375 |
1.00% |
0.25 |
||
5μL |
3.00% |
0.15 |
2.00% |
0.1 |
||
50-300μL |
5μL |
300μL |
0.70% |
2.1 |
0.25% |
0.75 |
150μL |
1.00% |
1.5 |
0.50% |
0.75 |
||
50μL |
1.50% |
0.75 |
0.80% |
0.4 |